Enhancing intraoral scanner accuracy using scan aid for multiple implants in the edentulous arch: an in vivo study

Abstract: Objectives
Intraoral scans of multiple implants in the edentulous arch are challenged by the absence of a distinct surface morphology between scan bodies. A scan aid was applied in such situation and evaluated for intraoral scanning accuracy in vivo.

Materials and Methods
87 implants in 22 patients were scanned with scan aid (SA) and without scan aid (NO) using two different intraoral scanners (CS3600 [CS] and TRIOS3 [TR]). Master casts were digitized by a laboratory scanner. Virtual models were superimposed using an inspection software and Linear deviation and precision were measured. Statistical analysis was performed using linear mixed models (α = .05).

Results
Total mean linear deviation within the CS group was 189 μm without scan aid and 135 μm when using the scan aid. The TR group's total mean deviation was 165 μm with and without a scan aid. Significant improvement with scan aid was observed for the CS group (p = .001), and no difference was found in the TR group. 96% of scan bodies were successfully scanned in the TR-SA group compared to 86% for the TR-NO group, 83% for the CS-SA, and 70% for the CS-NO group, respectively.

Conclusions
The evaluated scan aid improved linear deviation compared to unsplinted scans for the CS group but not for the TR group. These differences could originate from different scanning technologies used, active triangulation (CS) and confocal microscopy (TR). The scan aid improved the ability to recognize scan bodies successfully with both systems, which could have a favorable clinical impact overall

Standort
Deutsche Nationalbibliothek Frankfurt am Main
Umfang
Online-Ressource
Sprache
Englisch
Anmerkungen
Clinical oral implants research. - 34, 8 (2023) , 793-801, ISSN: 1600-0501

Ereignis
Veröffentlichung
(wo)
Freiburg
(wer)
Universität
(wann)
2023
Urheber

DOI
10.1111/clr.14107
URN
urn:nbn:de:bsz:25-freidok-2372778
Rechteinformation
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
Letzte Aktualisierung
25.03.2025, 13:52 MEZ

Datenpartner

Dieses Objekt wird bereitgestellt von:
Deutsche Nationalbibliothek. Bei Fragen zum Objekt wenden Sie sich bitte an den Datenpartner.

Beteiligte

Entstanden

  • 2023

Ähnliche Objekte (12)