Low dose pediatric chest computed tomography on a photon counting detector system – initial clinical experience

Abstract: Background

With the clinical release of a photon counting detector-based computed tomography (CT) system, the potential benefits of this new technology need to be evaluated clinically. Literature concerning this new generation of detector is sparse, especially in the field of pediatric radiology. Therefore, this study outlines our initial experience with ultra-low dose chest CT imaging on the new photon counting CT system.
Materials and methods

A pediatric phantom (1-year old, CIRS ATOM phantom, model 704 [CIRS-computerized imaging reference system, Norfolk, VA]) was scanned at different dose levels and different image quality levels to define a protocol for clinical examinations. Next, 20 consecutive pediatric non-contrast ultra-low dose chest CT examinations were evaluated for radiation dose and diagnostic image quality using a 4-point Likert-scale—1 = excellent, 4 = bad image quality—by two radiologists in a consensus reading. This retrospective analysis was approved by the local research ethics committee.
Results

Chest CT examinations performed at ultra-low radiation dose (effective dose 0.19 ± 0.07 mSv; size-specific dose estimate 0.45 ± 0.14 mGy) in pediatric patients ages (2.6 ± 1.8 years) show good to excellent image quality for lung structures (1.4 ± 0.4) and moderate image quality for soft tissue structures (2.8 ± 0.2).
Conclusion

Pediatric ultra-low dose chest CT examinations are feasible with the new generation photon counting detector-based CT system. The benefits of this technology must be evaluated for pediatric patients from the outset

Location
Deutsche Nationalbibliothek Frankfurt am Main
Extent
Online-Ressource
Language
Englisch
Notes
Pediatric radiology. - 53, 6 (2023) , 1057-1062, ISSN: 1432-1998

Event
Veröffentlichung
(where)
Freiburg
(who)
Universität
(when)
2023

DOI
10.1007/s00247-022-05584-4
URN
urn:nbn:de:bsz:25-freidok-2333201
Rights
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
Last update
25.03.2025, 1:49 PM CET

Data provider

This object is provided by:
Deutsche Nationalbibliothek. If you have any questions about the object, please contact the data provider.

Associated

Time of origin

  • 2023

Other Objects (12)