Pulsatile fatigue testing of arterial stents with radially applied load : An update based on new standardization

Abstract: Pulsatile fatigue testing is a basic requirement for arterial stents, however, there are several opportunities to setup and conduct the testing. The recently published standard guide ASTM F2477:2023 provides specific guidance, in particular includes the pressure controlled test method applying an outer hydraulic pressure to the test samples. The current work describes this approach, and demonstrates that the technical requirements can be met successfully. In addition, some inherent advantages are discussed, e.g. accuracy of pressure loading, absence of stent migration and potential high test frequencies.

Location
Deutsche Nationalbibliothek Frankfurt am Main
Extent
Online-Ressource
Language
Englisch

Bibliographic citation
Pulsatile fatigue testing of arterial stents with radially applied load ; volume:10 ; number:4 ; year:2024 ; pages:551-554 ; extent:4
Current directions in biomedical engineering ; 10, Heft 4 (2024), 551-554 (gesamt 4)

Creator
Schmidt, Wolfram
Brandt-Wunderlich, Christoph
Kamke, Frank
Schwerdt, Christopher
Kaminsky, Jörg
Behrens, Peter
Schmitz, Klaus-Peter
Grabow, Niels

DOI
10.1515/cdbme-2024-2135
URN
urn:nbn:de:101:1-2412181747390.903064022749
Rights
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
Last update
15.08.2025, 7:32 AM CEST

Data provider

This object is provided by:
Deutsche Nationalbibliothek. If you have any questions about the object, please contact the data provider.

Associated

  • Schmidt, Wolfram
  • Brandt-Wunderlich, Christoph
  • Kamke, Frank
  • Schwerdt, Christopher
  • Kaminsky, Jörg
  • Behrens, Peter
  • Schmitz, Klaus-Peter
  • Grabow, Niels

Other Objects (12)