Systematic morphological characterization of sub-visible particles generated from cardiovascular devices using dynamic image analysis

Abstract: The assessment of particulate matter generated during a simulated use is requested by international standards for approval of cardiovascular devices. The particle suspension was generated during a simulated use test procedure of a commercially available coronary stent delivery system and analyzed by Dynamic image analysis (DIA) using FlowCAM 8000. Four parameters were analyzed: the diameter, the aspect ratio value, the circularity and the intensity. The results revealed that 53% of the particles were larger than 25 μm. The particle shapes were inhomogeneous with a wide ranging aspect ratio of 0.03 to 0.95. Around 83% had an intensity value of more than 140 grey scales and appeared translucent.

Standort
Deutsche Nationalbibliothek Frankfurt am Main
Umfang
Online-Ressource
Sprache
Englisch

Erschienen in
Systematic morphological characterization of sub-visible particles generated from cardiovascular devices using dynamic image analysis ; volume:7 ; number:2 ; year:2021 ; pages:617-620 ; extent:4
Current directions in biomedical engineering ; 7, Heft 2 (2021), 617-620 (gesamt 4)

Urheber
Kurzhals, Anja
Brandt-Wunderlich, Christoph
Schmitz, Klaus-Peter
Grabow, Niels
Schmidt, Wolfram

DOI
10.1515/cdbme-2021-2157
URN
urn:nbn:de:101:1-2022101214153058207630
Rechteinformation
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
Letzte Aktualisierung
15.08.2025, 07:32 MESZ

Datenpartner

Dieses Objekt wird bereitgestellt von:
Deutsche Nationalbibliothek. Bei Fragen zum Objekt wenden Sie sich bitte an den Datenpartner.

Beteiligte

  • Kurzhals, Anja
  • Brandt-Wunderlich, Christoph
  • Schmitz, Klaus-Peter
  • Grabow, Niels
  • Schmidt, Wolfram

Ähnliche Objekte (12)