Lowering the radioactivity of the photomultiplier tubes for the XENON1T dark matter experiment

Abstract: The low-background, VUV-sensitive 3-inch diameter photomultiplier tube R11410 has been developed by Hamamatsu for dark matter direct detection experiments using liquid xenon as the target material. We present the results from the joint effort between the XENON collaboration and the Hamamatsu company to produce a highly radio-pure photosensor (version R11410-21) for the XENON1T dark matter experiment. After introducing the photosensor and its components, we show the methods and results of the radioactive contamination measurements of the individual materials employed in the photomultiplier production. We then discuss the adopted strategies to reduce the radioactivity of the various PMT versions. Finally, we detail the results from screening 286 tubes with ultra-low background germanium detectors, as well as their implications for the expected electronic and nuclear recoil background of the XENON1T experiment

Location
Deutsche Nationalbibliothek Frankfurt am Main
Extent
Online-Ressource
Language
Englisch
Notes
The European physical journal. C. - 75 (2015) , 546, ISSN: 1434-6052

Event
Veröffentlichung
(where)
Freiburg
(who)
Universität
(when)
2021
Creator
Aprile, Elena
Schumann, Marc
Laubenstein, Matthias
XENON Collaboration, [Study Group]

DOI
10.1140/epjc/s10052-015-3657-5
URN
urn:nbn:de:bsz:25-freidok-2200417
Rights
Kein Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
Last update
14.08.2025, 10:48 AM CEST

Data provider

This object is provided by:
Deutsche Nationalbibliothek. If you have any questions about the object, please contact the data provider.

Associated

Time of origin

  • 2021

Other Objects (12)