Mobile air quality monitoring and comparison to fixed monitoring sites for instrument performance assessment
Abstract ± 4 ppbv of the fixed reference site for nitrogen dioxide and ozone and up to about 8 ppbv for nitric oxide. We propose that this agreement can be leveraged to assess instrument performance over time during large-scale mobile monitoring campaigns. We demonstrate an example of how such relationships can be employed during large-scale monitoring campaigns using small sensors to identify potential measurement biases.
- Standort
-
Deutsche Nationalbibliothek Frankfurt am Main
- Umfang
-
Online-Ressource
- Sprache
-
Englisch
- Erschienen in
-
Mobile air quality monitoring and comparison to fixed monitoring sites for instrument performance assessment ; volume:17 ; number:9 ; year:2024 ; pages:2991-3009 ; extent:19
Atmospheric measurement techniques ; 17, Heft 9 (2024), 2991-3009 (gesamt 19)
- Urheber
-
Whitehill, Andrew R.
Lunden, Melissa
LaFranchi, Brian
Kaushik, Surender
Solomon, Paul A.
- DOI
-
10.5194/amt-17-2991-2024
- URN
-
urn:nbn:de:101:1-2405230425348.748602125189
- Rechteinformation
-
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
- Letzte Aktualisierung
-
14.08.2025, 10:52 MESZ
Datenpartner
Deutsche Nationalbibliothek. Bei Fragen zum Objekt wenden Sie sich bitte an den Datenpartner.
Beteiligte
- Whitehill, Andrew R.
- Lunden, Melissa
- LaFranchi, Brian
- Kaushik, Surender
- Solomon, Paul A.