The FreiBurger: a new optotype for P300-based acuity estimation

Abstract: Purpose
Accurate objective assessment of visual acuity is crucial, particularly in cases of suspected malingering, or when the patient’s inability to cooperate makes standard psychophysical acuity tests unreliable. The P300 component of the event-related potentials offers a potential solution and even allows for the use of standard optotypes like the Landolt C. However, low-vision patients with large eccentric visual field defects often struggle to locate the Landolt C gap quickly enough for a P300 to be reliably produced.

Methods
Addressing this challenge, we introduce a novel optotype (the “FreiBurger”) with a critical detail that extends through the optotype’s center. Two experiments, with 16 and 12 participants, respectively, were conducted. In the first, psychophysical acuity estimates were obtained with both the FreiBurger and the Landolt C. In the second, we tested the performance of the FreiBurger, relative to the Landolt C, in eliciting a P300 with undegraded vision, simulated low vision, and in a simulated combination of low vision and visual field constriction.

Results
Comparable psychophysical acuity values (average difference 0.03 logMAR) were obtained for both optotypes. In the P300 recordings, both optotypes produced similar P300 responses under conditions of undegraded vision and low vision. However, with the combination of low vision and constricted visual field, the P300 could only be reliably obtained with the FreiBurger, while the amplitude was drastically reduced with the Landolt C (9.1 µV vs. 2.2 µV; p < 0.0005).

Conclusion
The new optotype extends the applicability of P300-based acuity estimation to the frequently encountered combination of low vision and constricted visual field, where Landolt C optotypes fail. Although impairments were simulated in the present study, we assume that the advantages of the new optotype will also manifest in patients with such impairments. We furthermore expect the advantages to apply to time-sensitive psychophysical examinations as well

Location
Deutsche Nationalbibliothek Frankfurt am Main
Extent
Online-Ressource
Language
Englisch
Notes
Documenta ophthalmologica. - 149, 2 (2024) , 87-98, ISSN: 1573-2622

Event
Veröffentlichung
(where)
Freiburg
(who)
Universität
(when)
2024
Creator

DOI
10.1007/s10633-024-09982-x
URN
urn:nbn:de:bsz:25-freidok-2516413
Rights
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
Last update
15.08.2025, 7:24 AM CEST

Data provider

This object is provided by:
Deutsche Nationalbibliothek. If you have any questions about the object, please contact the data provider.

Associated

Time of origin

  • 2024

Other Objects (12)