CFD Investigation of Non‐Condensable Gases in Vacuum and Non‐Vacuum Steam Sterilizers
Abstract: Steam sterilization processes rely on high heat transfer rates from the steam to the medical devices. To guarantee this, all non‐condensable gases (NCGs) must be removed from the steam sterilizer before sterilization. The ratios and distribution of the NCGs in the pre‐sterilization phases of both a vacuum and a non‐vacuum steam sterilizer are investigated. Two CFD models for simulation of the NCGs in the sterilizers were developed and validated by the measurements of a pressure sensor and eight thermocouples inside the steam sterilizer. Furthermore, a novel method was developed to calculate the real ratio of NCGs based on the standard measurement (EN 13060). The results show that the NCGs ratio defined in the EN 13060 did not accurately represent the real ratios of the NCGs.
- Location
-
Deutsche Nationalbibliothek Frankfurt am Main
- Extent
-
Online-Ressource
- Language
-
Englisch
- Bibliographic citation
-
CFD Investigation of Non‐Condensable Gases in Vacuum and Non‐Vacuum Steam Sterilizers ; volume:91 ; number:4 ; year:2019 ; pages:502-513 ; extent:12
Chemie - Ingenieur - Technik ; 91, Heft 4 (2019), 502-513 (gesamt 12)
- Creator
-
Feurhuber, Manuel
Magno, Marino
Miranda, Marco
Prieler, René
Hochenauer, Christoph
- DOI
-
10.1002/cite.201800088
- URN
-
urn:nbn:de:101:1-2022072908323063981419
- Rights
-
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
- Last update
-
15.08.2025, 7:35 AM CEST
Data provider
Deutsche Nationalbibliothek. If you have any questions about the object, please contact the data provider.
Associated
- Feurhuber, Manuel
- Magno, Marino
- Miranda, Marco
- Prieler, René
- Hochenauer, Christoph