Reliable water vapour isotopic composition measurements at low humidity using frequency-stabilised cavity ring-down spectroscopy

Abstract ppmv, the instrument demonstrates a precision of 0.01 ‰ and 0.1 ‰ in δ 18 O and d-excess, respectively, for an integration time of 2 min. This set-up yields an isotopic composition precision below 1 ‰ at water mixing ratios down to 4 ppmv, which suggests an extrapolated precision in δ 18 O of 1.5 ‰ at 1 ppmv. Indeed, thanks to the stabilisation of the laser by the VCOF, the instrument exhibits extremely low drift and very high signal-to-noise ratio. The instrument is not hindered by a strong isotope–humidity response which at low humidity can create extensive biases on commercial instruments.

Standort
Deutsche Nationalbibliothek Frankfurt am Main
Umfang
Online-Ressource
Sprache
Englisch

Erschienen in
Reliable water vapour isotopic composition measurements at low humidity using frequency-stabilised cavity ring-down spectroscopy ; volume:17 ; number:15 ; year:2024 ; pages:4599-4612 ; extent:14
Atmospheric measurement techniques ; 17, Heft 15 (2024), 4599-4612 (gesamt 14)

Urheber
Casado, Mathieu
Landais, Amaelle
Stoltmann, Tim
Chaillot, Justin
Daëron, Mathieu
Prié, Fréderic
Bordet, Baptiste
Kassi, Samir

DOI
10.5194/amt-17-4599-2024
URN
urn:nbn:de:101:1-2408150438240.003665425924
Rechteinformation
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
Letzte Aktualisierung
14.08.2025, 10:50 MESZ

Datenpartner

Dieses Objekt wird bereitgestellt von:
Deutsche Nationalbibliothek. Bei Fragen zum Objekt wenden Sie sich bitte an den Datenpartner.

Beteiligte

  • Casado, Mathieu
  • Landais, Amaelle
  • Stoltmann, Tim
  • Chaillot, Justin
  • Daëron, Mathieu
  • Prié, Fréderic
  • Bordet, Baptiste
  • Kassi, Samir

Ähnliche Objekte (12)