Design and Characterisation of an EIT Voltage Conditioning Module

Abstract: Electrical Impedance Tomography is used to image the cross-sectional conductivity of an object. It is clinically used for high-frame-rate imaging of lung ventilation. Most current systems use very limited current injection waveforms and patterns. We have developed a flexible system for researching alternatives. This paper covers our voltage conditioning module’s design and characterisation. The device was shown to work up to the desired 10 MHz. The responses were uniform with minor gain mismatch between all channels after manufacturing variations. The low-frequency (<1 MHz) crosstalk is on the order -55 dB, and the worst cases around 8 MHz are around -35 dB.

Standort
Deutsche Nationalbibliothek Frankfurt am Main
Umfang
Online-Ressource
Sprache
Englisch

Erschienen in
Design and Characterisation of an EIT Voltage Conditioning Module ; volume:10 ; number:4 ; year:2024 ; pages:673-676 ; extent:4
Current directions in biomedical engineering ; 10, Heft 4 (2024), 673-676 (gesamt 4)

Urheber
Wilkie, Jack
Battistel, Alberto
Chen, Rongqing
Möller, Knut

DOI
10.1515/cdbme-2024-2165
URN
urn:nbn:de:101:1-2412181801538.776710429145