Development of an artifact-free aneurysm clip : Concept and approach for the development of an aneurysm clip entirely made from fiber-reinforced plastic materials

Abstract: For the treatment of intracranial aneurysms with aneurysm clips, usually a follow-up inspection in MRI is required. To avoid any artifacts, which can make a proper diagnosis difficult, a new approach for the manufacturing of an aneurysm clip entirely made from fiber-reinforced plastics has been developed. In this paper the concept for the design of the clip, the development of a new manufacturing technology for the fiber-reinforced components as well as first results from the examination of the components in phantom MRI testing is shown.

Location
Deutsche Nationalbibliothek Frankfurt am Main
Extent
Online-Ressource
Language
Englisch

Bibliographic citation
Development of an artifact-free aneurysm clip ; volume:2 ; number:1 ; year:2016 ; pages:543-546 ; extent:4
Current directions in biomedical engineering ; 2, Heft 1 (2016), 543-546 (gesamt 4)