Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems

Abstract: In order to demonstrate the influence of the boundary conditions in experimental biomechanical investigations of arthrodesis implants two different models were investigated. As basic model, a simplified finite element model of the cortical bone was used in order to compare the stress values with (Model 1) and without (Model 2) allowing horizontal displacements of the load application point. The model without constraints of horizontal displacements showed considerably higher stress values at the point of failure. Moreover, this investigation shows that the boundary conditions (e.g. constraints) have to be carefully considered, since simplifications of the reality with experimental tests cannot always be avoided.

Location
Deutsche Nationalbibliothek Frankfurt am Main
Extent
Online-Ressource
Language
Englisch

Bibliographic citation
Possible reasons for early artificial bone failure in biomechanical tests of ankle arthrodesis systems ; volume:1 ; number:1 ; year:2015 ; pages:507-509 ; extent:3
Current directions in biomedical engineering ; 1, Heft 1 (2015), 507-509 (gesamt 3)

Creator
Martin, H.
Gutteck, N.
Matthies, J.-B.
Hanke, T.
Gradl, G.
Wohlrab, D.
Mittlmeier, Th.
Grabow, N.

DOI
10.1515/cdbme-2015-0121
URN
urn:nbn:de:101:1-2410141643539.905452165880
Rights
Open Access; Der Zugriff auf das Objekt ist unbeschränkt möglich.
Last update
15.08.2025, 7:28 AM CEST

Data provider

This object is provided by:
Deutsche Nationalbibliothek. If you have any questions about the object, please contact the data provider.

Associated

  • Martin, H.
  • Gutteck, N.
  • Matthies, J.-B.
  • Hanke, T.
  • Gradl, G.
  • Wohlrab, D.
  • Mittlmeier, Th.
  • Grabow, N.

Other Objects (12)